Attaching Pro-Tip Turbo to Pro-Tip Converter
  • How do I attach Pro-Tip Turbo to Pro-Tip Converter?
    Whilst holding the converter outer sleeve scallops between thumb and index finger push the Pro-Tip firmly into the outer sleeve with one of the external ribs aligned with the dot on the outer sleeve surface. At this point twist the Pro-Tip to engage the locking feature.

  • Can I alter the air and water pressure?
    If necessary minor adjustments in air and water pressure can be made firstly at the dental unit and then at the junction box regulator.

  • Pro-Tip Turbo is too tight- what should I do?
    The Pro-Tip may appear dry and stiff to load at first. Place a small amount of Petroleum Jelly on the back face of the first Pro-Tip. This will lubricate the sealing 'O' rings. Repeat periodically to maintain lubrication.


Cleaning and Sterilising
  • How can I clean Pro-Tip Turbo and Pro-Tip Converters?
    All external surfaces of the Pro-Tip converter and the 3 in 1 syringe should be cleaned with a suitable mild detergent and water solution.Following removal of the converter from the syringe clean all internal surfaces of the converter and ‘O’ rings with a suitable mild detergent and water solution.

  • Can I sterilise Pro-Tip Converters?
    The Pro-Tip converter should be sterilised using a standard steam autoclave cycle to 134°C for a minimum of 3 minutes. Do not use dry heat sterilisation method.


Air and/or Water leakage

O Rings